Background:
Clinical outcome studies for cardiac troponins (cTn) are expensive and difficult to design owing to variation in patients, in the assays, and in the incidence of different types of myocardial infarction (MI). To overcome these difficulties, simulation models were used to estimate the rate of misclassification error for MI and risk prediction resulting from assay bias and imprecision.
Methods:
Finite mixture analysis of Abbott high-sensitivity cTnI (hs-cTnI) results at time 0 h in patients presenting early with acute coronary syndrome (ACS) symptoms to the emergency department (ED) [n = 145, Reducing the Time Interval for Identifying New Guideline (RING) study] allowed derivation of a simulation data set (n = 10000). hs-cTnI concentrations were modified by addition of bias or imprecision error. The percentage of all 10000 modified hs-cTnI results that were misclassified for MI at thresholds of 2, 5, 26.2, and 52 ng/L was determined by Monte Carlo simulation. Analyses were replicated with an all-comer emergency department (ED) population (n = 1137) ROMI (Optimum Troponin Cutoffs for ACS in the ED) study.
Results:
In the RING study, simulation at 26.2-ng/L (99th percentile) and 52-ng/L thresholds were affected by both bias ±2 ng/L and imprecision (10%-20%) and had misclassification rates of 0.4% to 0.6%. Simulations at the 2-ng/L and 5-ng/L thresholds were only affected by bias. Misclassification rates at bias of ±1 ng/L were 10% for the 2-ng/L threshold, and 5% for the 5-ng/L threshold.
Conclusions:
Simulation models predicted that hs-cTnI results are seldom misclassified (<1% of patients) when interpretative thresholds are near or exceed the overall 99th percentile. However, simulation models also predicted that low hs-cTnI results, as recommended in guidelines, are prone to misclassification of 5%-10% of patients.
More Related Videos
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
10:01Improvement of a Closed Chest Porcine Myocardial Infarction Model by Standardization of Tissue and Blood Sampling Procedures
Published on: March 12, 2018
Related Concept Videos
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Mechanistic Models: Compartment Models in Individual and Population Analysis
Receiver Operating Characteristic Plot
Errors occurring during blood pressure monitoring
Several factors...
Errors In Hypothesis Tests