Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy
Ata Mahmoodpoor1, Hadi Hamishehkar2, Kamran Shadvar1
1Department of Anesthesiology, Faculty of Medicine, Tabriz University of Medical Sciences, Tabriz, Iran.
Background And Aims:
There have been variable results on the practice of tight glycemic control, and studies have demonstrated that point-of-care (POC) glucometers have variable accuracy. Glucometers must be accurate, and many variables can affect blood glucose levels. The purpose of this study was to determine the difference between blood glucose concentrations obtained from POC glucometers and laboratory results in critically ill patients with intensive insulin therapy.
Materials And Methods:
This was a descriptive study which enrolled 300 critically ill patients. Four samples of arterial blood were collected and analyzed at the bedside with the POC glucometer and also in the central laboratory to obtain the blood glucose level. To define the effect of various factors on this relation, we noted the levels of hemoglobin (Hb), PaO2, body temperature, bilirubin, history of drug usage, and sepsis.
Results:
There were not any significant differences between blood sugar levels using laboratory and glucometer methods of measurements. There was a good and significant correlation between glucose levels between two methods (r = 0.81, P < 0.001). Among evaluated factors (body temperature, bilirubin level, blood pressure, Hb level, PaO2, sepsis, and drugs) which added one by one in model, just drugs decreased the correlation more than others (r = 0.78).
Conclusions:
The results of POC glucometer differ from laboratory glucose concentrations, especially in critically ill patients with unstable hemodynamic status while receiving several drugs. This may raise the concern about using POC devices for tight glycemic control in critically ill patients. These results should be interpreted with caution because of the large variation of accuracy among different glucometer devices.
More Related Videos
06:59Characterization of Metabolic Status in Nonhuman Primates with the Intravenous Glucose Tolerance Test
Published on: November 13, 2016
07:35Author Spotlight: Investigating the Blood Glucose Homeostasis in Murine Brain Using a Cost-Effective Hyperglycemic And Hypoglycemic Clamp Technique
Published on: January 26, 2024
Related Concept Videos
Hypoglycemia and Glucagon
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Diabetes: Symptoms, Diagnosis, and Complications
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...