Air bubbles are released by thoracic endograft deployment: An in vitro experimental study
Kamuran Inci1, Giasemi Koutouzi2, Valery Chernoray3
1Department of Surgery, Varberg Hospital, Varberg, Sweden.
SAGE Open Medicine
|December 21, 2016
Summary
Air bubbles are released during thoracic endograft deployment, potentially causing stroke after thoracic endovascular aortic repair. This study quantifies air release from two common endograft models.
Area of Science:
- Biomedical Engineering
- Vascular Surgery
- Cardiovascular Research
Background:
- Embolic stroke is a significant risk associated with thoracic endovascular aortic repair (TEVAR).
- Current understanding attributes TEVAR-related embolic stroke primarily to dislodged atherosclerotic debris.
- An alternative hypothesis suggests air embolism from the endograft delivery system as a potential cause.
Purpose of the Study:
- To experimentally determine if air is released during the deployment of thoracic endografts.
- To investigate air release as a potential alternative source of embolism during TEVAR.
Main Methods:
- Benchtop experimental study involving the deployment of eight thoracic endografts (Medtronic Valiant Thoracic and Gore TAG) in a de-aired, water-filled container.
- Endografts were prepared and deployed following manufacturer instructions.
- Released air volume was collected, measured using a calibrated syringe, and deployment was filmed.
Main Results:
- Air release was observed from all tested thoracic endografts.
- Quantified air volumes ranged from 0.1 to 0.3 mL for Medtronic Valiant Thoracic and <0.025 to 0.04 mL for Gore TAG.
- The largest observed bubbles, approximately 3 mm in diameter, originated from the proximal end of the Medtronic Valiant device.
Conclusions:
- Thoracic endografts release air bubbles upon deployment.
- Air embolism represents a plausible alternative mechanism for stroke occurrence following TEVAR.
- Further investigation into air management during TEVAR is warranted.