Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Ultrasound image based visual servoing for moving target ablation by high intensity focused ultrasound
Joonho Seo1, Norihiro Koizumi2, Mamoru Mitsuishi3
1Korea Institute of Machinery and Materials, Daegu, South Korea.
Background:
Although high intensity focused ultrasound (HIFU) is a promising technology for tumor treatment, a moving abdominal target is still a challenge in current HIFU systems. In particular, respiratory-induced organ motion can reduce the treatment efficiency and negatively influence the treatment result. In this research, we present: (1) a methodology for integration of ultrasound (US) image based visual servoing in a HIFU system; and (2) the experimental results obtained using the developed system.
Materials And Methods:
In the visual servoing system, target motion is monitored by biplane US imaging and tracked in real time (40 Hz) by registration with a preoperative 3D model. The distance between the target and the current HIFU focal position is calculated in every US frame and a three-axis robot physically compensates for differences. Because simultaneous HIFU irradiation disturbs US target imaging, a sophisticated interlacing strategy was constructed.
Results:
In the experiments, respiratory-induced organ motion was simulated in a water tank with a linear actuator and kidney-shaped phantom model. Motion compensation with HIFU irradiation was applied to the moving phantom model. Based on the experimental results, visual servoing exhibited a motion compensation accuracy of 1.7 mm (RMS) on average. Moreover, the integrated system could make a spherical HIFU-ablated lesion in the desired position of the respiratory-moving phantom model.
Conclusions:
We have demonstrated the feasibility of our US image based visual servoing technique in a HIFU system for moving target treatment.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
08:02A High-Throughput Image-Guided Stereotactic Neuronavigation and Focused Ultrasound System for Blood-Brain Barrier Opening in Rodents
Published on: July 16, 2020
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography