Related Experiment Video
Updated: Mar 9, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Suction catheter for enhanced control and accuracy of transseptal access
Matthew S Sulkin1, Zachary C Berwick, James B Hermiller
13DT Holdings LLC, San Diego, CA, USA.
Aims:
Percutaneous structural heart therapies, such as mitral value repair, require site-specific transseptal access (TSA). This can be challenging for interventional cardiologists. We describe a TSA catheter (TSAC) that utilises suction for enhanced control and puncture accuracy. Here, we aim to evaluate the safety and efficacy of the device.
Methods And Results:
Ex vivo interatrial septum preparations were dissected from swine (n=8) and diseased human hearts (n=6) to quantify TSAC suction and needle puncture force. TSAC suction was 6.5-fold greater than the opposing needle puncture force, and thus provides sufficient stabilisation for punctures. The safety and efficacy of TSAC was evaluated in a chronic mitral regurgitation swine model (n=10) and compared to a conventional TSA device. MR was induced by disrupting one to three mitral chordae tendineae, and the progression of heart disease was followed for three weeks. During device testing, procedure time and fluoroscopy exposure were not statistically different between devices. TSAC reduced septal displacement from 8.7±0.30 mm to 3.60±0.19 mm (p<0.05) and improved puncture accuracy 1.75-fold.
Conclusions:
TSAC provides controlled TSA and improves puncture accuracy, while maintaining procedure time and workflow. These findings provide a strong rationale for a first-in-man study to demonstrate the clinical utility of the device.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
07:47The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
Published on: December 23, 2014
Related Concept Videos
Tracheostomy Suctioning II: Procedure
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...