A Feasibility Study for Measuring Accurate Chest Compression Depth and Rate on Soft Surfaces Using Two Accelerometers
Sofía Ruiz de Gauna1, Digna M González-Otero1, Jesus Ruiz1
1Department of Communications Engineering, University of the Basque Country (UPV/EHU), Alameda Urquijo S/N, 48013 Bilbao, Spain.
A new algorithm using two accelerometers accurately measures cardiopulmonary resuscitation (CPR) compression depth and rate, even on soft surfaces. This system corrects for mattress displacement, preventing overestimation of depth during resuscitation efforts.
Area of Science:
- Biomedical Engineering
- Emergency Medicine
- Cardiopulmonary Resuscitation Technology
Background:
- Accelerometer-based cardiopulmonary resuscitation (CPR) feedback devices are common.
- Existing devices overestimate chest compression depth on soft surfaces, potentially leading to ineffective resuscitation.
- Accurate depth measurement is crucial for effective CPR.
Purpose of the Study:
- To evaluate a novel algorithm for CPR compression depth and rate measurement using dual accelerometers.
- To assess the algorithm's performance in a simulated resuscitation environment with varying surfaces.
- To address the limitations of current devices in soft surface scenarios.
Main Methods:
- A manikin was used for compressions on foam and sprung mattresses, with and without a backboard.
- Two accelerometers were placed on the manikin's chest and back.
- A new algorithm calculated sternal-spinal displacement by subtracting mattress displacement from total chest displacement, derived from spectral analysis of acceleration data.
Main Results:
- The system demonstrated a median unsigned error of 2.1 mm (4.4%) for compression depth.
- Depth measurement errors varied between mattress types (foam: 2.4 mm, sprung: 1.7 mm) and backboard use.
- Median error for compression rate was 0.9 cpm (1.0%), with no significant differences across conditions.
Conclusions:
- The dual-accelerometer system provides accurate CPR feedback on compression depth and rate, even on soft surfaces.
- The algorithm effectively compensates for mattress displacement, overcoming a key limitation of current devices.
- This technology can improve the quality of CPR delivered in challenging environments.
Related Concept Videos
Physical Assessment of the Respiratory Tract III: Percussion
Percussion in Respiratory Assessment
Percussion evaluates underlying tissue composition with audible and tactile vibrations,...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: