Manual resuscitators in successive use in the same patient: reservoir of multi- and extensively resistant bacteria
G Pinheiro Lima Aires Gomes1, A Custódia Silva E Souza2, L S Netto de Oliveira Leão-Vasconcelos3
1Faculty of Nursing, Federal University of Goias, Goiânia, Brazil.
Abstract:
To determine the bacterial load on manual resuscitators (MRs), swab samples of MR connectors from a Brazilian intensive care unit were collected at time zero, and after 24 and 48h use in the same patient. From 20 MRs, a total of 54 bacteria were isolated: six at t=0, 17 at t=24, and 31 at t=48. Staphylococcus aureus, Acinetobacter baumannii, Serratia marcescens, and Pseudomonas aeruginosa were the predominant isolates, and multidrug resistance was widespread in all species. Greater awareness of the need to ensure effective decontamination of MRs between patients, as well as more frequent replacement of MRs in individual patients, is recommended to minimize the risk of (re)colonization/infection of the respiratory tract.
Insights
Bacterial contamination significantly increases on manual resuscitators (MRs) during patient use. Effective decontamination and more frequent MR replacement are crucial to prevent respiratory tract infections.
Area of Science:
- Infectious Diseases
- Microbiology
- Critical Care Medicine
Background:
- Manual resuscitators (MRs) are critical for ventilation in intensive care units (ICUs).
- The potential for bacterial contamination of MRs poses a risk to patient health.
- Standard decontamination protocols may not be sufficient to eliminate microbial load.
Purpose of the Study:
- To quantify bacterial load on manual resuscitators (MRs) during use in an ICU.
- To identify predominant bacterial species and assess their antimicrobial resistance.
- To inform recommendations for MR handling and patient safety.
Main Methods:
- Swab samples collected from MR connectors of 20 devices at 0, 24, and 48 hours of use in the same patient.
- Bacterial isolation and identification using standard microbiological techniques.
- Antimicrobial susceptibility testing to determine multidrug resistance patterns.
Main Results:
- A total of 54 bacterial isolates were recovered from 20 MRs.
- Bacterial load increased significantly over time: 6 isolates at 0h, 17 at 24h, and 31 at 48h.
- Predominant isolates included Staphylococcus aureus, Acinetobacter baumannii, Serratia marcescens, and Pseudomonas aeruginosa, with widespread multidrug resistance.
Conclusions:
- Manual resuscitators accumulate significant bacterial contamination during patient use.
- Multidrug-resistant bacteria commonly colonize MRs, increasing infection risk.
- Enhanced decontamination and more frequent MR replacement are recommended to mitigate patient (re)colonization and infection risk.