Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
Published on: April 29, 2011
Photoplethysmography for an independent measure of pulsatile pressure under controlled flow conditions
Haneen Njoum1, Panayiotis A Kyriacou1
1Research Centre for Biomedical Engineering, City University London, London, EC1V 0HB, UK.
Abstract:
Noninvasive continuous blood pressure measurements are desirable for patients and clinicians. This work proposes and validates a method for transmural pressure measurement using photoplethysmography (PPG) in an in vitro setup that allows control of pressure and flow conditions. The optimum pulsatile volume measure is obtained by comparing parameters extracted from the photoplethysmographic signal (AC amplitude, normalized pulse volume (NPV) and adjusted pulse volume (APV)). Pulsatile volume can then provide pressure measurements using the exponential pressure-volume (P-V) relationship and validated using the gold standard catheter pressure measurement. Pressure, red (R) and infrared (IR) PPG signals were recorded continuously in two arterial models with different cross-sectional areas (Model 1 and Model 2) utilising a pulsatile pump. Flow rates were controlled by varying pumping frequencies at low and high stroke volumes. The optimum method for estimation of the pulsatile volume is through APV, which had a highly significant correlation (r 2 = 0.99, p < 0.001) for Model 1 and (r 2 = 0.98, p < 0.001) for Model 2. APV obtained a significantly better fit when compared to NPVIR (r 2 = 0.73, z = 25.85, p < 0.001), NPVR (r 2 = 0.95, z = 12.26, p < 0.001), IRAC (r 2 = 0.52, z = 28.29, p < 0.0001) and RAC (r 2 = 0.92, z = 15.27, p < 0.0001) in Model 1, and when compared to NPVIR (r 2 = 0.92, z = 10.23, p < 0.0001), NPVR (r 2 = 0.96, z = 5.08, p < 0.001) IRAC (r 2 = 0.63, z = 22.47, p < 0.0001) and RAC (r 2 = 0.92, z = 17.70, p < 0.0001) in Model 2. These preliminary findings suggest that APV could be used as a potential non-invasive continuous method of blood pressure measurement at different flow conditions.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...
Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)