Related Experiment Video
Updated: Mar 9, 2026

Diagnosing Pulmonary Tuberculosis with the Xpert MTB/RIF Test
Published on: April 9, 2012
Performance of Xpert MTB/RIF on Ascitic Fluid Samples for Detection of Abdominal Tuberculosis
Syed Beenish Rufai1, Sarman Singh1, Amit Singh1
1Division of Clinical Microbiology and Molecular Medicine, Department of Laboratory Medicine, All India Institute of Medical Sciences, New Delhi, India.
Background:
Diagnosis of abdominal tuberculosis (TB) from ascitic fluid samples using routinely available diagnostic methods is challenging due to its paucibacillary nature. Although performance of Xpert MTB/RIF assay has been evaluated extensively on pulmonary samples, its performance on extrapulmonary samples is still under evaluation.
Objectives:
The objective of this study was to find out the performance of Xpert MTB/RIF on ascitic fluid samples obtained from suspected cases of abdominal TB. Performance was compared with Mycobacterium growth indicator tube-960 (MGIT-960) culture and in-house multiplex polymerase chain reaction (PCR). The latter detects and differentiates Mycobacterium tuberculosis and nontuberculous mycobacteria simultaneously.
Materials And Methods:
Sixty-seven patients suspected of probable/possible abdominal TB were included in this observational, prospective study. All samples were tested by Ziehl-Neelsen staining, MGIT-960 culture, in-house multiplex PCR, and Xpert MTB/RIF assay.
Results:
All 67 samples were smear negative. Seventeen (25.4%) were MGIT-960 culture positive while 12 (17.9%) were detected positive by the Xpert MTB/RIF assay and 9 (13.4%) by in-house multiplex PCR. Sensitivity and specificity of the Xpert MTB/RIF assay compared with the MGIT-960 culture were 70.6% (95%, confidence interval [CI]: 44.1-89.7) and 100% (95%, CI: 92.8-100) and that of in-house multiplex PCR were 52.9% (95%, CI: 30.9-73.8) and 100% (95%, CI: 92.8-100), respectively.
Conclusions:
Diagnostic yield of Xpert MTB/RIF assay on ascitic fluid samples was lower than MGIT-960 culture. We thus emphasize on the need for urgent discovery of new biomarkers for paucibacillary TB.
More Related Videos
09:54Synthesis, Characterization, and Application of Superparamagnetic Iron Oxide Nanoprobes for Extrapulmonary Tuberculosis Detection
Published on: February 16, 2020
03:47A Retrospective Study on Endoscopic Surgery for the Treatment of Paravertebral Abscess in Spinal Tuberculosis Patients
Published on: October 25, 2024
Related Concept Videos
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...
Pulmonary Tuberculosis III
The first classification is based on the development of the disease, and it includes the following categories: