Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Reply to: Comment on: polarity effects and apparent ion recombination in microionization chambers [Med. Phys. 43(5)
Jessica R Miller1, Brian D Hooten2, John A Micka3
1Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, 53792, USA.
Abstract:
We would like to thank Dr. Brivio et al. [Med. Phys.] for their comment on our recent paper. Miller et al. [Med. Phys. 43 (2016) 2141-2152] determined the primary cause of voltage-dependent polarity effects in microchambers to be a potential difference between the guard and collecting electrodes. In their comment, Brivio et al., offer an explanation for the cause of such potential differences. Brivio et al. attribute the potential difference to the disparity in the work functions between guard and collecting electrodes composed of different materials. However, all of the microchambers investigated in Miller et al. contained a guard and collecting electrode which were composed of the same material. Therefore, the explanation offered by Brivio et al. that "the electric potential perturbation arises from the work function difference of the disparate materials electrodes" does not explain the polarity effects exhibited by the microchambers investigated in Miller et al., all of which contain electrodes composed of the same materials.
Related Concept Videos
Ion-Exchange Chromatography
Ion Exchange
Mass Analyzers: Common Types
Chemical Ionization (CI) Mass Spectrometry
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Potential Due to a Polarized Object