Assessment of chest CT at CTDIvol less than 1 mGy with iterative reconstruction techniques
Atul Padole1, Subba Digumarthy1, Efren Flores1
11 Department of Radiology, Massachusetts General Hospital, Boston, MA, USA.
Objective:
To assess the image quality of chest CT reconstructed with image-based iterative reconstruction (SafeCT; MedicVision®, Tirat Carmel, Israel), adaptive statistical iterative reconstruction (ASIR; GE Healthcare, Waukesha, WI) and model-based iterative reconstruction (MBIR; GE Healthcare, Waukesha, WI) techniques at CT dose index volume (CTDIvol) <1 mGy.
Methods:
In an institutional review board-approved study, 25 patients gave written informed consent for acquisition of three reduced dose (0.25-, 0.4- and 0.8-mGy) chest CT after standard of care CT (8 mGy) on a 64-channel multidetector CT (MDCT) and reconstructed with SafeCT, ASIR and MBIR. Two board-certified thoracic radiologists evaluated images from the lowest to the highest dose of the reduced dose CT series and subsequently for standard of care CT.
Results:
Out of the 182 detected lesions, the missed lesions were 35 at 0.25, 24 at 0.4 and 9 at 0.8 mGy with SafeCT, ASIR and MBIR, respectively. The most missed lesions were non-calcified lung nodules (NCLNs) 25/112 (<5 mm) at 0.25, 18/112 (<5 mm) at 0.4 and 3/112 (<4 mm) at 0.8 mGy. There were 78%, 84% and 97% lung nodules detected at 0.25, 0.4 and 0.8 mGy, respectively regardless of iterative reconstruction techniques (IRTs), Most mediastinum structures were not sufficiently seen at 0.25-0.8 mGy.
Conclusion:
NCLNs can be missed in chest CT at CTDIvol of <1 mGy (0.25, 0.4 and 0.8 mGy) regardless of IRTs. The most lung nodules (97%) were detected at CTDIvol of 0.8 mGy. The most mediastinum structures were not sufficiently seen at 0.25-0.8 mGy. Advances in knowledge: NCLNs can be missed regardless of IRTs in chest CT at CTDIvol of <1 mGy. The performance of ASIR, SafeCT and MBIR was similar for lung nodule detection at 0.25, 0.4 and 0.8 mGy.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...