Effect of axial temperature gradient on chromatographic efficiency under adiabatic conditions
Krisztián Horváth1, Szabolcs Horváth1, Diána Lukács1
1Department of Analytical Chemistry, University of Pannonia, Egyetem utca 10, H-8200 Veszprém, Hungary.
Abstract:
The effect of axial temperature gradient on the chromatographic efficiency was studied under adiabatic conditions by a modeling approach. The equilibrium-dispersive model of chromatography was used for the calculations. The model was extended by taking into account the axial temperature gradient. The results show that due to the temperature gradient, there are retention and migration velocity gradients in the column. Since the retention factor, k, is not constant in the column, k cannot be calculated as the ratio of net retention and hold-up times. As a result of the gradual increase of migration velocity, the retention times of solutes decrease as the slope of temperature gradient increases. In addition, the band in the column have extra broadening due to larger migration velocity of the front of band. The width of bands becomes larger at larger change of temperature. In the same time, however, the release velocity of the compounds from the column is increasing as ΔT increases. Accordingly, an apparent peak compression effect makes the peaks thinner. As a result of the two counteracting effects (peak expansion, apparent peak compression) the column efficiency does not change significantly in case of axial temperature gradient under adiabatic conditions. The resolutions, however, decrease slightly due to the decrease of retention times.
Related Concept Videos
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Principles Of Column Chromatography
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...