Bone marrow biopsies performed by both the powered OnControl drill device and the Jamshidi needle produce adequate

Sarika Jain1, Mark Enzerra2, Rohtesh S Mehta2

  • 1Department of Pathology, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, USA.

Abstract

Insights

Bone marrow biopsy adequacy was comparable between the manual Jamshidi needle and the OnControl powered drill. However, the OnControl device resulted in a higher aspiration artefact rate.

Area of Science:

  • Hematology
  • Medical Devices
  • Surgical Techniques

Background:

  • Bone marrow biopsy is crucial for diagnosing hematological disorders.
  • Traditional manual bone marrow biopsy methods can be invasive and may yield suboptimal specimens.
  • Powered drills offer potential for improved bone marrow specimen acquisition.

Purpose of the Study:

  • To compare the adequacy and quality of bone marrow specimens obtained using the OnControl powered drill versus a manual Jamshidi needle.
  • To evaluate specific parameters such as biopsy length, evaluable marrow area, and artefacts.

Main Methods:

  • Retrospective analysis of 75 bone marrow specimens.
  • Comparison between specimens obtained by Jamshidi needle (n=44) and OnControl powered drill (n=31).
  • Evaluation of biopsy length, evaluable marrow length and area, and artefacts (fragmentation, aspiration, marrow dropout).

Main Results:

  • Adequacy for diagnosis was similar: 86% for manual (M-group) and 83% for powered drill (P-group).
  • Common inadequacy reasons: subcortical specimens (M-group) and aspiration artefact (P-group).
  • Aspiration artefact was significantly higher in the OnControl group (p<0.0001).

Conclusions:

  • Bone marrow biopsy adequacy and quality are comparable between manual and powered drill methods.
  • The OnControl powered drill is associated with a higher incidence of aspiration artefact.
  • Both methods provide sufficient bone marrow specimens for diagnosis, with differing artefact profiles.