Impact of Systemic Vascular Resistance on the Accuracy of the Pulsioflex Device
Matthieu Biais1, Elodie Mazocky, Laurent Stecken
1From the *CHU de Bordeaux, Service d'Anesthésie Réanimation 3, Bordeaux, France; †INSERM and ‡University Bordeaux, Adaptation cardiovasculaire à l'ischémie, U1034, Pessac, France; §CHU de Bordeaux, Service d'Anesthésie réanimation 1, Bordeaux, France; and ‖Biostatistics Unit (DRCI), Délégation Recherche Clinique and Innovation, CHU de Clermont-Ferrand, Villa annexe IFSI, Rue Montalembert, Clermont-Ferrand Cedex, France.
Background:
The accuracy of currently available devices using pulse contour analysis without external calibration for cardiac index (CI) estimation is negatively impacted by hyperdynamic states, low systemic vascular resistance (SVR), and abrupt changes in SVR. The aim of this study was to evaluate the accuracy of a new device, the Pulsioflex (Pulsion Medical System), in patients undergoing liver transplantation.
Methods:
Thirty consecutive patients scheduled for liver transplantation were included. CI was monitored using pulmonary arterial catheter (CI-PAC) and Pulsioflex (CI-Pulsio). Simultaneous CI measurements were made intraoperatively at 9 different stages of the procedure.
Results:
Two hundred seventy pairs of measurements were analyzed. The median CI-Pulsio values (3.3; interquartile range, 2.8-3.8 L·min·m) were significantly different from the median CI-PAC (4.1; interquartile range, 3.1-5.0 L·min·m; P < .0001). Bland and Altman analysis showed a mean bias of 0.8 L·min·m and 95% limit of agreement from -2.5 to 4.1 L·min·m. Percentage error was 65% (95% confidence interval, 60%-71%). Considering the variations in CI between 2 stages, the comparison between changes in CI-PAC and changes in CI-Pulsio showed a mean bias of 0.1 L·min·m and 95% limit of agreement of -2.1 to 2.2 L·min·m. When excluding changes in CI <0.5 L·min·m (154 paired analyzed), the concordance rate was 62% (95% confidence interval, 54%-70%). The bias between CI-PAC and CI-Pulsio was negatively correlated with SVR (r = -0.67, P < .0001). The bias between changes in CI-PAC and changes in CI-Pulsio was also negatively correlated with changes in SVR (r = -0.52, P < .0001).
Conclusions:
In patients undergoing liver transplantation, Pulsioflex does not accurately estimate CI. Its accuracy is highly impacted by SVR, and it is not able to track changes in CI when large variations in SVR occur.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Vascular Resistance
The primary determinants of vascular resistance are vessel diameter, blood viscosity, and vessel length. Among these, vessel diameter plays the most significant role due to the fourth power relationship described by...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...