Clinical Implementation of Dual-energy CT for Proton Treatment Planning on Pseudo-monoenergetic CT scans
Patrick Wohlfahrt1, Christian Möhler2, Volker Hietschold3
1OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden - Rossendorf, Dresden, Germany; Helmholtz-Zentrum Dresden - Rossendorf, Institute of Radiooncology, Dresden, Germany.
Purpose:
To determine whether a standardized clinical application of dual-energy computed tomography (DECT) for proton treatment planning based on pseudomonoenergetic CT scans (MonoCTs) is feasible and increases the precision of proton therapy in comparison with single-energy CT (SECT).
Methods And Materials:
To define an optimized DECT protocol, CT scan settings were analyzed experimentally concerning beam hardening, image quality, and influence on the heuristic conversion of CT numbers into stopping-power ratios (SPRs) and were compared with SECT scans with identical CT dose. Differences in range prediction and dose distribution between SECT and MonoCT were quantified for phantoms and a patient.
Results:
Dose distributions planned on SECT and MonoCT datasets revealed mean range deviations of 0.3 mm, γ passing rates (1%, 1 mm) greater than 99.9%, and no clinically relevant changes in dose-volume histograms. However, image noise and CT-related uncertainties could be reduced by MonoCT compared with SECT, which resulted in a slightly decreased dependence of SPR prediction on beam hardening. Consequently, DECT was clinically implemented at the University Proton Therapy Dresden in 2015. Until October 2016, 150 patients were treated based on MonoCTs, and more than 950 DECT scans of 351 patients were acquired during radiation therapy.
Conclusions:
A standardized clinical use of MonoCT for treatment planning is feasible, leads to improved image quality and SPR prediction, extends diagnostic variety, and enables a stepwise clinical implementation of DECT toward a physics-based, patient-specific, nonheuristic SPR determination. Further reductions of CT-related uncertainties, as expected from such SPR approaches, can be evaluated on the resulting DECT patient database.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT