Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
Micro-electromechanical film bulk acoustic sensor for plasma and whole blood coagulation monitoring
Da Chen1, Shuren Song2, Jilong Ma2
1Mine Disaster Prevention and Control Key Laboratory of Education Ministry of China, Shandong University of Science and Technology, Qingdao 266590, China; College of Electronics, Communications, and Physics, Shandong University of Science and Technology, Qingdao 266590, PR China.
Abstract:
Monitoring blood coagulation is an important issue in the surgeries and the treatment of cardiovascular diseases. In this work, we reported a novel strategy for the blood coagulation monitoring based on a micro-electromechanical film bulk acoustic resonator. The resonator was excited by a lateral electric field and operated under the shear mode with a frequency of 1.9GHz. According to the apparent step-ladder curves of the frequency response to the change of blood viscoelasticity, the coagulation time (prothrombin time) and the coagulation kinetics were measured with the sample consumption of only 1μl. The procoagulant activity of thromboplastin and the anticoagulant effect of heparin on the blood coagulation process were illustrated exemplarily. The measured prothrombin times showed a good linear correlation with R2=0.99969 and a consistency with the coefficient of variation less than 5% compared with the commercial coagulometer. The proposed film bulk acoustic sensor, which has the advantages of small size, light weight, low cost, simple operation and little sample consumption, is a promising device for miniaturized, online and automated analytical system for routine diagnostics of hemostatic status and personal health monitoring.