Should patients with higher blood pressure variability be excluded from validation studies? An assessment of the
Raj S Padwal1, Donna McLean, Jennifer Ringrose
1aDepartment of Medicine bFaculty of Nursing, University of Alberta cAlberta Diabetes Institute dMazankowski Heart Institute, Edmonton, Alberta, Canada.
Abstract:
To limit the inclusion of participants with increased blood pressure (BP) variability and presumably to avoid potential bias, the International Standards Organization BP device validation standard recommends exclusion of patients with a BP variability of more than 12/8 mmHg across reference readings. This '12/8 rule' is based on expert consensus and lacks empirical justification. In a post-hoc analysis of a study comparing two types of cuff designs carried out according to the International Standards Organization standard, we divided the study sample into patients who did not have (n=79) and patients who had (n=55) more than 12/8 mmHg variability. Patients with more than 12/8 mmHg variability were older and had a higher prevalence of diabetes (41.8 vs. 22.8%; P=0.02) and hypertension (43.6 vs. 29.1%; P=0.08). The mean systolic BP differences between the two cuff designs were not significantly different in participants who did not show more than 12/8 mmHg variability versus those who did (2.2±3.5 vs. 3.1±3.4; mean difference of differences -0.9±3.4; P=0.14). Similarly, the mean diastolic BP differences were not significantly different in participants who did not have more than 12/8 mmHg variability versus those who did (1.5±2.2 vs. 1.4±2.6; mean difference of differences 0.1±2.4; P=0.82). A limitation of our analysis is that the original study data focused on a comparison of different cuff designs and not formal validation of a specific device. Therefore, replication of these findings is warranted. Nevertheless, our findings do not support the use of the 12/8 rule and indicate that this rule may be promoting unnecessarily homogenous study samples, limiting external generalizability, and needlessly increasing workload and expense.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: