Locking Cap Designs Improve Fatigue Properties of Polyaxial Screws in Upper Extremity Applications
Surena Namdari1, Samir Mehta, Ann Tierney
1*Biedermann Lab for Orthopaedic Research, Department of Orthopaedic Surgery, University of Pennsylvania, Philadelphia, PA; †Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, PA; and ‡Center for Clinical Epidemiology and Biostatistics, Department of Biostatistics and Epidemiology, University of Pennsylvania, Philadelphia, PA.
Objectives:
This study sought to examine fatigue characteristics of 2 polyaxial locking screw designs: locking cap (LC) and cross-threaded (CT). The goal was to compare LC and CT implants at 0, 10, and 15 degrees of angulation to determine the effect of locking mechanism on screw-plate interface failure. The hypothesis was that LC implants would have superior fatigue properties in comparison to CT designs and that increased angulation of the screw would have a negative impact on the fatigue life of CT implants, but would not have any effect on LC implants.
Methods:
A total of 72 screws were tested in 4 upper extremity implants. Implants were subjected to cyclic shear loads and subsequent ramp to failure. Performance characteristics were statistically compared using nonparametric statistical methods.
Results:
Fatigue testing demonstrated that LC designs were consistently able to sustain a significantly higher number of cyclic loads than CT designs. There were no significant differences in the number of cycles sustained by LC designs because of changes in screw angle, but CT implants exhibited decreases in screw stability with increasing angulation.
Conclusions:
Likely because of the spherical screw head geometry, LC fatigue characteristics are not influenced by the orientation of the screw relative to the plate. Application of an LC in the operating room requires additional time, but provides significantly more robust fixation of the screw, especially at oblique angles to the plate and provides a more predictable and consistent biomechanical result.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
12:25C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Circular Shafts - Elastoplastic Materials
As torque on the...
Stress Concentrations in Circular Shafts
Residual Stresses in Circular Shafts