Purpose:
The goal of this study is to characterize and improve the accuracy of 2D magnetic resonance fingerprinting (MRF) scans in the presence of slice profile (SP) and B1 imperfections, which are two main factors that affect quantitative results in MRF.
Methods:
The SP and B1 imperfections are characterized and corrected separately. The SP effect is corrected by simulating the radiofrequency pulse in the dictionary, and the B1 is corrected by acquiring a B1 map using the Bloch-Siegert method before each scan. The accuracy, precision, and repeatability of the proposed method are evaluated in phantom studies. The effects of both SP and B1 imperfections are also illustrated and corrected in the in vivo studies.
Results:
The SP and B1 corrections improve the accuracy of the T1 and T2 values, independent of the shape of the radiofrequency pulse. The T1 and T2 values obtained from different excitation patterns become more consistent after corrections, which leads to an improvement of the robustness of the MRF design.
Conclusion:
This study demonstrates that MRF is sensitive to both SP and B1 effects, and that corrections can be made to improve the accuracy of MRF with only a 2-s increase in acquisition time. Magn Reson Med 78:1781-1789, 2017. © 2017 International Society for Magnetic Resonance in Medicine.
More Related Videos
07:42Optimizing Magnetic Force Microscopy Resolution and Sensitivity to Visualize Nanoscale Magnetic Domains
Published on: July 20, 2022
10:28Measuring Interactions of Globular and Filamentous Proteins by Nuclear Magnetic Resonance Spectroscopy NMR and Microscale Thermophoresis MST
Published on: November 2, 2018
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
Interpreting ¹H NMR Signal Splitting: The (n + 1) Rule
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
¹H NMR Signal Multiplicity: Splitting Patterns