Implementation and application of moving average as continuous analytical quality control instrument demonstrated for
Background:
General application of a moving average (MA) as continuous analytical quality control (QC) for routine chemistry assays has failed due to lack of a simple method that allows optimization of MAs. A new method was applied to optimize the MA for routine chemistry and was evaluated in daily practice as continuous analytical QC instrument.
Methods:
MA procedures were optimized using an MA bias detection simulation procedure. Optimization was graphically supported by bias detection curves. Next, all optimal MA procedures that contributed to the quality assurance were run for 100 consecutive days and MA alarms generated during working hours were investigated.
Results:
Optimized MA procedures were applied for 24 chemistry assays. During this evaluation, 303,871 MA values and 76 MA alarms were generated. Of all alarms, 54 (71%) were generated during office hours. Of these, 41 were further investigated and were caused by ion selective electrode (ISE) failure (1), calibration failure not detected by QC due to improper QC settings (1), possible bias (significant difference with the other analyzer) (10), non-human materials analyzed (2), extreme result(s) of a single patient (2), pre-analytical error (1), no cause identified (20), and no conclusion possible (4).
Conclusions:
MA was implemented in daily practice as a continuous QC instrument for 24 routine chemistry assays. In our setup when an MA alarm required follow-up, a manageable number of MA alarms was generated that resulted in valuable MA alarms. For the management of MA alarms, several applications/requirements in the MA management software will simplify the use of MA procedures.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include:
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Development of Analytical Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Quality Assurance