[Diagnostic accuracy of outpatient polygraphy devices : A comparison with inpatient polysomnography in clinical
J Lindemann1, B Augenstein2, F Stupp2
1Klinik für Hals-, Nasen- und Ohrenheilkunde, Kopf- und Halschirurgie, Universitätsklinikum Ulm, Frauensteige 12, 89075, Ulm, Deutschland. joerg.lindemann@uniklinik-ulm.de.
Background:
Cardiorespiratory polysomnography (PSG) is considered the reference method for diagnosis of obstructive sleep apnea (OSA). Due to waiting times and high costs, payers increasingly request outpatient polygraphy (PG) as an alternative to inpatient PSG. The aim of the present study was to evaluate the diagnostic accuracy of different outpatient PG devices compared to stationary PSG in clinical practice.
Materials And Methods:
Externally collected outpatient PG findings of 406 patients were retrospectively compared with the corresponding PSG findings.
Results:
Among the 406 patients were 343 men (85%) and 63 women (15%), with mean age 50 years. Mean body mass index (BMI) was 30 kg/m2. The rank correlation coefficient for PG- and PSG- derived apnea-hypopnea index (AHI) values was r = 0.574. On average, PG underestimated the AHI by 6.4 (±20.5) events/h. OSAS severity was determined correctly by PG in only 43% of cases. Sensitivity (90.7%) and specificity (45.2%) of ambulatory PG was calculated for the threshold value AHI ≥ 5/h. Based on the results of PG, an indicated therapy would have been omitted in 35 cases (9%) and unnecessary treatment initiated in 17 cases (4%). The PG devices used showed a comparable diagnostic accuracy (r = 0.513-0.657), with a sensitivity of 81.3-96.9% and a specificity of 33.3-50.0%.
Conclusion:
Outpatient PG cannot reliably assess OSA severity in clinical routine. Confirmation by PSG in a sleep lab in symptomatic patients is obligatory. Outpatient PG devices should only be used as an upstream screening method. The automatic evaluation of the PG should always be proofed.
More Related Videos
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
04:33Author Spotlight: Unveiling the Connection Between Sleep Disorders and Cognitive Symptoms in Depression
Published on: April 26, 2024
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring
Receiver Operating Characteristic Plot