Comparison of Two Ambient Beta Gauge PM10 Samplers
Chuen JinnTsai, Yu-Hsiang Cheng1
1a Institute of Environmental Engineering, National Chiao Tung University , Hsin-Chu , Taiwan , USA.
Abstract:
This paper reports on the performance of the Kimoto 180 sampler and the Wedding ambient PM10 beta gauge sampler. Monodisperse ammonium fluorescein test particles were generated in the laboratory and used to determine the penetration curve of the Kimoto 180 cyclonic inlet. It was found that the actual DpaSOof the Kimoto 180 inlet, 3.5 (xm, is much lower than the designated standard value, 10 um. In the field test, the two beta gauge samplers were collocated with an Andersen SA1200 high-volume sampler to compare their measured daily average PM10 concentrations. The low Dpa50 of the Kimoto 180 inlet serves to explain why its daily average PM10 concentrations were much lower than the actual PM10 concentrations found in the field study. In addition, the PM10 concentrations of the Kimoto 180 beta gauge sampler were found to be seriously affected by the water vapor content of the ambient air. In contrast, the daily average PM10 concentrations of the Wedding beta gauge sampler were found to be more accurate, and influences by ambient conditions were insignificant.
More Related Videos
09:33Visualizing Field Data Collection Procedures of Exposure and Biomarker Assessments for the Household Air Pollution Intervention Network Trial in India
Published on: December 23, 2022
10:29Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
Related Concept Videos
Pressure Gauges
Measurement of Air Content in Concrete
The pressure method,...