Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in
Kerry L Tomlin1, Anna-Maria Neitenbach1, Ulf Borg2
1Department of Medical Affairs, Patient Monitoring & Recovery, Medtronic, 6135 Gunbarrel Ave, Boulder, CO, 80301, USA.
Background:
Regional oximetry is increasingly used to monitor post-extraction oxygen status of the brain during surgical procedures where hemodynamic fluctuations are expected. Particularly in cardiac surgery, clinicians employ an interventional algorithm to restore baseline regional oxygen saturation (rSO2) when a patient reaches a critical desaturation threshold. Evidence suggests that monitoring cardiac surgery patients and intervening to maintain rSO2 can improve postoperative outcomes; however, evidence generated with one manufacturer's device may not be applicable to others. We hypothesized that regional oximeters from different manufacturers respond uniquely to changes in oxygen saturation in healthy volunteers.
Methods:
Three devices were tested: INVOS™ 5100C (Medtronic), EQUANOX™ 7600 (Nonin), and FORE-SIGHT™ (CASMED) monitors. We divided ten healthy subjects into two cohorts wearing a single sensor each from INVOS and EQUANOX (n = 6), or INVOS and FORE-SIGHT (n = 4). We induced and reversed hypoxia by adjusting the fraction of inspired oxygen. We calculated the magnitude of absolute rSO2 change and rate of rSO2 change during desaturation and resaturation, and determined if and when each device reached a critical interventional rSO2 threshold during hypoxia.
Results:
All devices responded to changes in oxygen directionally as expected. The median absolute rSO2 change and the rate of rSO2 change was significantly greater during desaturation and resaturation for INVOS compared with EQUANOX (P = 0.04). A similar but nonsignificant trend was observed for INVOS compared with FORE-SIGHT; our study was underpowered to definitively conclude there was no difference. A 10% relative decrease in rSO2 during desaturation was detected by all three devices across the ten subjects. INVOS met a 20% relative decrease threshold in all subjects of both cohorts, compared to 1 with EQUANOX and 2 with FORE-SIGHT. Neither EQUANOX nor FORE-SIGHT reached a 50% absolute rSO2 threshold compared with 4 and 3 subjects in each cohort with INVOS, respectively.
Conclusions:
Significant differences exist between the devices in how they respond to changes in oxygen saturation in healthy volunteers. We suggest caution when applying evidence generated with one manufacturer's device to all devices.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...