Image noise reduction technology reduces radiation in a radial-first cardiac catheterization laboratory
Ateka Gunja1, Yagya Pandey1, Hui Xie2
1Department of Veterans Affairs, Jesse Brown VA Medical Center, Chicago, IL; Division of Cardiology, Department of Medicine, University of Illinois at Chicago, Chicago, IL.
Background:
Transradial coronary angiography (TRA) has been associated with increased radiation doses. We hypothesized that contemporary image noise reduction technology would reduce radiation doses in the cardiac catheterization laboratory in a typical clinical setting.
Methods And Results:
We performed a single-center, retrospective analysis of 400 consecutive patients who underwent diagnostic and interventional cardiac catheterizations in a predominantly TRA laboratory with traditional fluoroscopy (N=200) and a new image noise reduction fluoroscopy system (N=200). The primary endpoint was radiation dose (mGy cm2). Secondary endpoints were contrast dose, fluoroscopy times, number of cineangiograms, and radiation dose by operator between the two study periods. Radiation was reduced by 44.7% between the old and new cardiac catheterization laboratory (75.8mGycm2±74.0 vs. 41.9mGycm2±40.7, p<0.0001). Radiation was reduced for both diagnostic procedures (45.9%, p<0.0001) and interventional procedures (37.7%, p<0.0001). There was no statistically significant difference in radiation dose between individual operators (p=0.84). In multivariate analysis, radiation dose remained significantly decreased with the use of the new system (p<0.0001) and was associated with weight (p<0.0001), previous coronary artery bypass grafting (p<0.0007) and greater than 3 stents used (p<0.0004). TRA was used in 90% of all cases in both periods. Compared with a transfemoral approach (TFA), TRA was not associated with higher radiation doses (p=0.20).
Conclusions:
Image noise reduction technology significantly reduces radiation dose in a contemporary radial-first cardiac catheterization clinical practice.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization III: Left Heart Catheterization
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...