Related Experiment Video
Updated: Feb 19, 2026

Optimization of Antigen Preparation for Hemagglutination Inhibition Assay of Newcastle Disease Virus Serology
Published on: January 24, 2025
Comparison of potential higher order reference methods for total haemoglobin quantification-an interlaboratory study
Claudia Frank1, Christine Brauckmann2, Marie Palos3
1Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, Germany. Claudia.Frank@ptb.de.
Abstract:
The total haemoglobin (Hb) concentration in blood is one of the most frequently measured analytes in clinical medicine because of its significance for evaluating the health state of a human. The spectrophotometric cyanmethaemoglobin (HiCN) method is the internationally accepted conventional reference method to determine this biomarker. It is frequently used in clinical routine diagnostics but is not traceable to the International System of Units and thus does not meet highest metrological demands. A further critical issue is the toxicity of the necessary potassium cyanide. Different methods to solve these problems are reported here. They all were validated against the HiCN method in an interlaboratory comparison by measuring the total Hb concentration present in the certified reference material JCCRM 912-2M. Methods considered were the spectrophotometric alkaline haematin detergent (AHD) method as well as several isotope dilution (ID)-based approaches. The latter include inductively coupled plasma mass spectrometry (ICP-MS), species-specific (SS) ICP-MS, organic MS and Raman spectrometry. Graphical abstract ᅟ.

