Study Objectives:
Home sleep apnea testing (HSAT) is increasingly available for diagnosing obstructive sleep apnea (OSA). One key limitation of most HSAT involves the lack of sleep staging, such that the respiratory event index is calculated using the total recording time (TRT) rather than total sleep time (TST).
Methods:
We performed a retrospective analysis of n = 838 diagnostic polysomnography (PSG) nights from our center; n = 444 with OSA (4% rule, apneahypopnea index (AHI) ≥ 5), and n = 394 with AHI < 5. We recalculated the AHI using time in bed (TIB) instead of TST, to assess the predicted underestimation risk of OSA severity.
Results:
Of all the patients with OSA, 26.4% would be reclassified as having less severe or no OSA after recalculating the AHI using TIB rather than TST. Of the n = 275 with mild OSA, 18.5% would be reclassified as not having OSA. The risk of underestimation was higher in those with moderate or severe OSA. Of the n = 119 moderate OSA cases, 40.3% would be reclassified as mild, and of the n = 50 severe OSA cases, 36.0% would be reclassified as moderate. Age strongly correlated with the degree of underestimation of the AHI, because age was significantly correlated with time awake during PSG.
Conclusions:
The risk of sleep apnea underestimation is predicted to be substantial in a tertiary sleep center population. Phenotyping errors included risk of falsely negative results (from mild to normal), as well as category errors: moderate or severe moving to mild or moderate severity, respectively. Clinicians should recognize this underestimation limitation, which directly affects diagnostic phenotyping and thus therapeutic decisions.
Commentary:
A commentary on this article appears in this issue on page 531.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Errors occurring during blood pressure monitoring
Several factors...