Repeatability of whole-cornea measurements using an anterior segment imaging device based on OCT and Placido-disk
Noelia Martínez-Albert1, José J Esteve-Taboada1, Robert Montés-Micó1
1a Optometry Research Group (GIO), Department of Optics and Optometry and Vision Sciences , University of Valencia , Burjassot , Spain.
Background:
Current advances in refractive surgery require high accuracy of corneal measurements. We aim to evaluate the repeatability of topographic measurements using an integrated Placido disk - OCT device.
Methods:
Thirty right eyes of thirty healthy and young subjects were included. The topographic parameters of simulated keratometry, asphericity, minimum pachymetry, white-to-white distance, spherical aberration and corneal elevation were studied. Three measurements were taken by the same observer using the Visante omni device. The coefficient of repeatability (CoR), coefficient of variation (CoV) and intraclass correlation coefficient (ICC) were calculated to assess the repeatability. The Bland-Altman graphical method was also applied.
Results:
No significant differences between repeated measurements were found. The mean difference for simulated keratometry was between 0.01 and 0.02 D. For asphericity was -0.01 ± 0.02, and for minimum pachymetry, spherical aberration and WTW was equal or less than 0.02 mm. The mean differences for corneal elevation, in millimeters and diopters, were between 0.00 and 0.01. Overall, most topographic parameters had a CoV less than 0.4%, a CoR less than 0.3 and an ICC higher than 0.98.
Conclusions:
The Visante omni device provides good repeatability for topographical measurements of simulated keratometry, asphericity, pachymetry, spherical aberration and corneal elevation.