Validation of a smartphone-based point-of-care hemoglobin assay for use in dogs
Kate S Farrell1, Phillip L Chapman2, Lauren A Sullivan1
1Department of Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences.
Objective:
To validate a smartphone-based spectrophotometric assay for point-of-care (POC) measurement of hemoglobin (Hgb) concentration against the standard measurement method in dogs without increases in serum lipemia, hemolysis, or bilirubin.
Design:
Prospective observational study.
Setting:
University teaching hospital.
Animals:
One hundred thirty-nine dogs that had a CBC and corresponding biochemical profile submitted to the clinical pathology laboratory.
Interventions:
None.
Measurements And Main Results:
Blood tubes submitted for CBC were collected for hemoglobin (Hgb) measurements performed on a POC smartphone device. Each whole blood sample was run on the smartphone in duplicate using 2 strips, for a total of 4 POC Hgb readings per dog. Data collected for each dog included CBC Hgb value, 4 POC Hgb values, and select biochemistry values (bilirubin, lipemia, hemolysis, icterus). A calibration equation was estimated using a weighted linear regression: estimated CBC = (avgPOC-0.4871)/1.0015. For each dog, the percent error was computed between estimated and actual Hgb values; 95% of the percent errors ranged from -13.2% to 20.1%. The standard deviation of percent errors was 7.9% overall. When samples were further divided according to CBC Hgb concentration (low, normal or high), the standard deviation of percent error was 6.7% when Hgb<13 g/dL [130 g/L], 8.9% when Hgb 13-20 g/dL [130-200 g/L], and 6.5% when Hgb>20 g/dL [200 g/L]. The coefficient of variability among the 4 individual POC readings was 3.4%.
Conclusions:
A smartphone-based photometric method for measuring Hgb represents a clinically useful POC alternative to a standard laboratory Hgb measurement. There was excellent intrasample reproducibility, and the standard deviation of percent errors was relatively constant across CBC ranges. Additional sampling of patients with a greater range of diseases and biochemical abnormalities that may influence spectrophotometric assays (ie, abnormal bilirubin, icterus, lipemia, hemolysis) is warranted to extend the findings of this study.