Related Experiment Video
Updated: Mar 8, 2026

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Systolic peak foot-to-apex time interval, a novel oscillometric technique for systolic blood pressure measurement
Amir M Benmira1, Antonia Perez-Martin, Sarah Coudray
1aVascular Medicine and Laboratory, Nîmes University Hospital, Nîmes, France bBiomedical Engineering Laboratory, Faculty of Technology, Aboubekr Belkaid University, Tlemcen, Algeria cResearch Unit (Female characteristics of dysfunctions of vascular interfaces), Montpellier University, Montpellier, France.
Background:
Noninvasive blood pressure (BP) measurement is essential for the study of human physiology but automatic oscillometric devices only estimate SBP and DBP using various, undisclosed algorithms, precluding standardization and interchangeability. We propose a novel approach by tracking, during pneumatic cuff deflation, the time interval from the foot to the apex of the systolic peak of the oscillometric signal, which reaches a maximum concomitant with the first Korotkoff sound.
Method:
In 145 study participants and patients (group 1), we measured the systolic brachial artery blood pressure by Korotkoff sound recording, conventional oscillometry, and our fully automated systolic peak foot-to-apex time interval (SFATI) technique. In 35 other patients (group 2), we compared SFATI with intra-arterial measurement.
Results:
In group 1, the concordance correlation coefficient was 0.989 and 0.984 between SFATI and Korotkoff sounds, 0.884 and 0.917 between oscillometry and Korotkoff sounds, and 0.882 and 0.919 between SFATI and oscillometry, respectively, on the left and right arm. In group 2, it was 0.72 between SFATI and intra-arterial measurement, 0.67 between oscillometry and intra-arterial measurement, and 0.92 between SFATI and Korotkoff sounds. In 40 study participants, the reproducibility study yielded a concordance coefficient of 0.95 for SFATI and 0.94 for Korotkoff sounds.
Conclusion:
SFATI BP measurement shows an excellent concordance with the auscultatory technique, offering a major improvement over current oscillometric techniques and allowing standardization.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Assessing Blood pressure in the Leg
Preparation: