Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Comparison of different techniques of central venous pressure measurement in mechanically ventilated critically ill
C Roger1,2,3, L Muller1,2, B Riou4
1Division of Anesthesia Intensive Care Pain and Emergency, Nîmes University Hospital, Place de Professeur Robert Debré, Nîmes, 30029, France.
Background:
Several techniques exist for measuring central venous pressure (CVP) but little information is available about the accuracy of each method. The aim of this study was to compare different methods of CVP measurements in mechanically ventilated patients.
Methods:
CVP was measured in mechanically ventilated patients without spontaneous breathing using four different techniques: 1) end expiratory CVP measurement at the base of the" c" wave (CVPMEASURED), chosen as the reference method; 2) CVP measurement from the monitor averaging CVP over the cardiac and respiratory cycles (CVPMONITOR); 3) CVP measurement after a transient withdrawing of mechanical ventilation (CVPNADIR); 4) CVP measurement corrected for the transmitted respiratory pressure induced by intrinsic PEEP (calculated CVP: CVPCALCULATED). Bias, precision, limits of agreement, and proportions of outliers (difference > 2 mm Hg) were determined.
Results:
Among 61 included patients, 103 CVP assessments were performed. CVPMONITOR bias [-0.87 (1.06) mm Hg] was significantly different from those of CVPCALCULATED [1.42 (1.07), P < 0.001 and CVPNADIR (1.04 (1.29), P < 0.001]. The limits of agreement of CVPMONITOR [-2.96 to 1.21 mm Hg] were not significantly different to those of CVPNADIR (-1.49 to 3.57 mm Hg, P = 0.39) and CVPCALCULATED (-0.68 to 3.53 mm Hg, P = 0.31). The proportion of outliers was not significantly different between CVPMONITOR (n = 5, 5%) and CVPNADIR (n = 9, 9%, P = 0.27) but was greater with CVPCALCULATED (n = 16, 15%, P = 0.01).
Conclusions:
In mechanically ventilated patients, CVPMONITOR is a reliable method for assessing CVPMEASURED Taking into account transmitted respiratory pressures, CVPCALCULATED had a higher proportion of outliers and precision than CVPNADIR.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement