An Objective and Child-friendly Assessment of Arm Function by Using a 3-D Sensor
Published on: February 12, 2018
Upper extremity movement reliability and validity of the Kinect version 2
Lydia R Reither1, Matthew H Foreman1, Nicole Migotsky1
1a Program in Occupational Therapy , Washington University School of Medicine, Human Performance Laboratory , Saint Louis , MO , USA.
Marker-less motion detection systems, like the Kinect 1 and Kinect 2, show good reliability for within-day upper extremity motion analysis. However, between-day comparisons require caution due to movement variability.
Area of Science:
- Biomedical Engineering
- Rehabilitation Technology
- Motion Analysis
Background:
- Marker-less motion detection systems offer potential for clinical applications.
- Previous studies show varied accuracy for Kinect 2 in balance and joint localization.
- Limited research exists on Kinect 2's validity and reliability for upper extremity movements.
Purpose of the Study:
- To compare the reliability and validity of Kinect 1 and Kinect 2 against a video motion capture (VMC) system.
- To assess the suitability of Kinect systems for upper extremity motion analysis in a clinical setting.
Main Methods:
- One healthy adult male performed six upper extremity movements.
- Simultaneous data collection using Kinect 1, Kinect 2, and VMC systems.
- Data analysis using MATLAB, Microsoft Excel, and SPSS.
Main Results:
- Both Kinect systems demonstrated good within-day reliability.
- Between-day reliability was inconclusive due to movement repetition challenges.
- Kinect systems showed high correlation in motion patterns with VMC, despite differing range of motion magnitudes.
Conclusions:
- Kinect systems can be clinically useful for motion analysis with an understanding of their limitations.
- Within-day assessments are acceptable; between-day comparisons need cautious interpretation.
- Kinect 2 offers a cost-effective, portable, and user-friendly alternative to VMC.