Related Experiment Video
Updated: Mar 8, 2026

A Volumetric Method for Quantification of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage
Published on: July 28, 2018
Simulation of MR angiography imaging for validation of cerebral arteries segmentation algorithms
Artur Klepaczko1, Piotr Szczypiński1, Andreas Deistung2
1Institute of Electronics, Lodz University of Technology, Lodz, Poland.
Background And Objective:
Accurate vessel segmentation of magnetic resonance angiography (MRA) images is essential for computer-aided diagnosis of cerebrovascular diseases such as stenosis or aneurysm. The ability of a segmentation algorithm to correctly reproduce the geometry of the arterial system should be expressed quantitatively and observer-independently to ensure objectivism of the evaluation.
Methods:
This paper introduces a methodology for validating vessel segmentation algorithms using a custom-designed MRA simulation framework. For this purpose, a realistic reference model of an intracranial arterial tree was developed based on a real Time-of-Flight (TOF) MRA data set. With this specific geometry blood flow was simulated and a series of TOF images was synthesized using various acquisition protocol parameters and signal-to-noise ratios. The synthesized arterial tree was then reconstructed using a level-set segmentation algorithm available in the Vascular Modeling Toolkit (VMTK). Moreover, to present versatile application of the proposed methodology, validation was also performed for two alternative techniques: a multi-scale vessel enhancement filter and the Chan-Vese variant of the level-set-based approach, as implemented in the Insight Segmentation and Registration Toolkit (ITK). The segmentation results were compared against the reference model.
Results:
The accuracy in determining the vessels centerline courses was very high for each tested segmentation algorithm (mean error rate = 5.6% if using VMTK). However, the estimated radii exhibited deviations from ground truth values with mean error rates ranging from 7% up to 79%, depending on the vessel size, image acquisition and segmentation method.
Conclusions:
We demonstrated the practical application of the designed MRA simulator as a reliable tool for quantitative validation of MRA image processing algorithms that provides objective, reproducible results and is observer independent.
More Related Videos
05:23Author Spotlight: Noninvasive Cerebral Blood Flow Determination in Human Functional Brain Region for Diagnosis of Neurological Disorders
Published on: May 31, 2024
12:50Lesion Explorer: A Video-guided, Standardized Protocol for Accurate and Reliable MRI-derived Volumetrics in Alzheimer's Disease and Normal Elderly
Published on: April 14, 2014