Optimization of Couch Modeling in the Change of Dose Calculation Methods and Their Versions
Junichi Kuwahara1, Manabu Nakata, Takahiro Fujimoto
1Division of Clinical Radiology Service, Kyoto University Hospital.
Abstract:
In external radiotherapy, the X-ray beam passes through the treatment couch, leading to the dose reduction by the attenuation of the couch. As a method to compensate for the reduction, radiation treatment planning systems (RTPS) support virtual couch function, namely "couch modeling method". In the couch modeling method, the computed tomography (CT) numbers assigned to each structure should be optimized by comparing calculations to measurements for accurate dose calculation. Thus, re-optimization of CT numbers will be required when the dose calculation algorithm or their version changes. The purpose of this study is to evaluate the calculation accuracy of the couch modeling method in different calculation algorithms and their versions. The optimal CT numbers were determined by minimizing the difference between measured transmission factors and calculated ones. When CT numbers optimized by Anisotropic Analytical Algorithm (AAA) Ver. 8.6 were used, the maximum and the mean difference of transmission factor were 5.8% and 1.5%, respectively, for Acuros XB (AXB) Ver. 11.0. However, when CT numbers optimized by AXB Ver. 11.0 were used, they were 2.6% and 0.6%, respectively. The CT numbers for couch structures should be optimized when changing dose calculation algorithms and their versions. From the comparison of the measured transmission to calculation, it was found that the CT numbers had high accuracy.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Dosage Regimen Designs: Nomograms and Tabulations
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Dosage Regimen: Fixed Dose
Fixed-dose regimens can be used for various routes of administration, including intravenous (IV) injections and oral medications. For IV administration, a predetermined amount of the drug is...
Dosage Regimens: Designs and Approaches