Development and validation of LC-MS/MS with in-source collision-induced dissociation for the quantification of

Monique Zangarini1, Neil Rajan2, Marina Danilenko2

  • 1Newcastle Cancer Centre Pharmacology Group, Newcastle University, Newcastle upon Tyne, NE2 4HH, UK.

Bioanalysis
|January 24, 2017
PubMed
Abstract

Insights

A new HPLC-MS/MS method accurately quantifies pegcantratinib in skin tumor biopsies. This validated assay is crucial for assessing anticancer drug penetration in patients during clinical trials.

Area of Science:

  • Pharmacology
  • Analytical Chemistry
  • Oncology

Background:

  • Pegcantratinib is a mini-PEGylated K252a derivative investigated as an anticancer agent.
  • It functions by inhibiting tropomyosin receptor kinase (TRK).
  • Accurate quantification of drug levels in tumor tissue is essential for efficacy studies.

Purpose of the Study:

  • To develop and validate a sensitive bioanalytical method.
  • Quantify pegcantratinib concentrations in human skin tumor biopsies.
  • Determine tumor drug penetration for clinical evaluation.

Main Methods:

  • High-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS).
  • In-source collision-induced dissociation (CID) for enhanced sensitivity.
  • Method validation including linearity, precision, accuracy, recovery, and matrix effect.

Main Results:

  • A sensitive and specific HPLC-MS/MS method was developed and validated.
  • The assay demonstrated excellent precision (CV ≤8.5%) and accuracy (95-102%).
  • Linearity was established from 1-500 ng/ml with a limit of quantitation of 2.5 ng/ml.

Conclusions:

  • A robust method for pegcantratinib analysis in human tumor biopsies was successfully developed.
  • The validated method was effectively applied to analyze clinical trial samples.
  • This assay supports the clinical development of pegcantratinib as an anticancer therapy.