Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

3.8K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.8K
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

2.7K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.7K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

1.9K
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.9K
Special considerations while measuring blood pressure01:28

Special considerations while measuring blood pressure

1.4K
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
1.4K
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

1.3K
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
1.3K
Sites for measuring blood pressure01:21

Sites for measuring blood pressure

4.0K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
4.0K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The 2026 global roadmap for textile-integrated wearable technologies in health.

Physiological measurement·2026
Same author

SLCs in diabetic kidney disease: From pathogenic roles to therapeutic promises.

Life sciences·2026
Same author

Left atrium incisure curvature: an indicator for predicting atrial fibrillation-related cardioembolism.

BMC cardiovascular disorders·2026
Same author

One-dimensional well-defined peapod Co<sub>x</sub>Ni<sub>y</sub>@C nanocomposite hollow nanorod towards enhanced electromagnetic performance.

Advanced composites and hybrid materials·2026
Same author

Enhancing few-shot personalized cuffless blood pressure estimation with self-supervised learning.

Physiological measurement·2026
Same author

Self-supervised Learning with Demographic Information for Cuffless Blood Pressure Estimation.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference·2025

Related Experiment Video

Updated: Mar 8, 2026

Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
11:04

Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism

Published on: September 1, 2014

11.6K

Validation of an Adaptive Transfer Function Method to Estimate the Aortic Pressure Waveform.

Yang Yao, Lisheng Xu, Yingxian Sun

    IEEE Journal of Biomedical and Health Informatics
    |January 24, 2017
    PubMed
    Summary

    This study introduces an adaptive transfer function (ATF) method to accurately estimate the aortic pulse wave from the brachial pulse wave. The ATF method improves cardiovascular assessment by providing a reliable, noninvasive approach for analyzing aortic pulse wave characteristics.

    More Related Videos

    Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound
    10:08

    Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound

    Published on: December 2, 2014

    16.6K
    Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
    11:26

    Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression

    Published on: December 10, 2014

    12.9K

    Related Experiment Videos

    Last Updated: Mar 8, 2026

    Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
    11:04

    Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism

    Published on: September 1, 2014

    11.6K
    Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound
    10:08

    Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound

    Published on: December 2, 2014

    16.6K
    Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
    11:26

    Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression

    Published on: December 10, 2014

    12.9K

    Area of Science:

    • Cardiovascular Physiology
    • Biomedical Engineering
    • Signal Processing

    Background:

    • Aortic pulse wave analysis is crucial for cardiovascular assessment but challenging noninvasively.
    • Peripheral pulse wave estimation offers a potential solution for indirect aortic pulse wave measurement.

    Purpose of the Study:

    • To develop and validate an adaptive transfer function (ATF) method for estimating aortic pulse wave from brachial pulse wave.
    • To assess the accuracy of the ATF method in predicting key aortic pulse wave parameters.

    Main Methods:

    • Derived aortic and brachial pulse waves from 26 patients undergoing cardiac catheterization.
    • Developed generalized transfer functions (GTF) using an autoregressive exogenous model.
    • Adapted GTF by peak resonance frequency, determined individually using brachial systolic blood pressure.
    • Validated the ATF method using leave-one-out cross-validation.

    Main Results:

    • The ATF method demonstrated superior performance in estimating aortic pulse wave compared to previous methods.
    • Prediction errors for aortic systolic blood pressure and pulse pressure were minimal (0.2 ± 3.1 and -0.9 ± 3.1 mmHg).
    • Accurate prediction of augmentation index, notch amplitude, and ejection duration was achieved.

    Conclusions:

    • The adaptive transfer function (ATF) method provides a reliable noninvasive technique for estimating aortic pulse wave from brachial pulse wave.
    • This method enhances cardiovascular status assessment and prediction of key hemodynamic parameters.
    • The ATF method shows significant potential for clinical application in cardiovascular monitoring.