Related Experiment Video
Updated: Mar 8, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Long-term survival of Stanmore total hip replacements inserted with radiolucent bone cement
J A Wimhurst1, J L Hobby, C P Roberts
1Ipswich Hospital NHS Trust, Ipswich - UK.
Abstract:
Radio-opacifiers in bone cements are an accepted part of every-day practice. They have, however, been shown to be a potential cause of an increase in third body wear and to excite bone resorption in in vitro and in vivo studies. We reviewed the results of 228 consecutive Stanmore total hip replacements performed between 1981 and 1985 in 211 patients. All were inserted with radiolucent bone cement. Information regarding whether the prosthesis had been revised was available for all patients. Seventy-three patients (83 hips) were still alive and 41 patients (44 hips) were sufficiently healthy to attend clinic. Information regarding pain level was obtained from the remaining 32 patients. When revision of the implant was taken as the end-point, there was 95% ten-year survival, 91% fifteen-year survival and 75% eighteen-year survival. These long-term results of Stanmore THRs, performed in a district general hospital, with radiolucent bone cement, compare favourably with the other published series for this implant. We did not find the inability to see the bone cement a particular disadvantage when reviewing radiographs for signs of loosening. (Hip International 2002; 12: 274-80).