A novel cuffless device for self-measurement of blood pressure: concept, performance and clinical validation
N Boubouchairopoulou1, A Kollias1, B Chiu2
1Hypertension Center STRIDE-7, Third Department of Medicine, National and Kapodistrian University of Athens, School of Medicine, Sotiria Hospital, Athens, Greece.
Abstract:
A pocket-size cuffless electronic device for self-measurement of blood pressure (BP) has been developed (Freescan, Maisense Inc., Zhubei, Taiwan). The device estimates BP within 10 s using three embedded electrodes and one force sensor that is applied over the radial pulse to evaluate the pulse wave. Before use, basic anthropometric characteristics are recorded on the device, and individualized initial calibration is required based on a standard BP measurement performed using an upper-arm BP monitor. The device performance in providing valid BP readings was evaluated in 313 normotensive and hypertensive adults in three study phases during which the device sensor was upgraded. A formal validation study of a prototype device against mercury sphygmomanometer was performed according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 2013 protocol. The test device succeeded in obtaining a valid BP measurement (three successful readings within up to five attempts) in 55-72% of the participants, which reached 87% with device sensor upgrade. For the validation study, 125 adults were recruited and 85 met the protocol requirements for inclusion. The mean device-observers BP difference was 3.2±6.7 (s.d.) mm Hg for systolic and 2.6±4.6 mm Hg for diastolic BP (criterion 1). The estimated s.d. (inter-subject variability) were 5.83 and 4.17 mm Hg respectively (criterion 2). These data suggest that this prototype cuffless BP monitor provides valid self-measurements in the vast majority of adults, and satisfies the BP measurement accuracy criteria of the ANSI/AAMI/ISO 2013 validation protocol.
More Related Videos
05:51Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)