Fission 99Mo/99mTc Generators--a study of their quality
Nuklearmedizin. Nuclear Medicine
|October 1, 1989
Summary
Quality control of molybdenum-99/technetium-99m (99Mo/99mTc) generators showed satisfying results. Eluates demonstrated high radionuclidic and radiochemical purity, confirming reliable 99mTc-pertechnetate production for medical applications.
Area of Science:
- Nuclear Medicine
- Radiopharmacy
- Analytical Chemistry
Background:
- The widespread use of 99mTc necessitates stringent quality control of its primary generator systems.
- Fission-produced molybdenum-99 (99Mo) is the parent radionuclide in the most common 99mTc generators.
- Ensuring the purity and performance of these generators is critical for safe and effective diagnostic imaging.
Purpose of the Study:
- To assess the quality of seven commercially available 99Mo/99mTc generators in Germany.
- To evaluate the radionuclidic and radiochemical purity of 99mTc eluates.
- To determine the 99Mo breakthrough, elution efficiency, pH, and aluminum content of the eluates, alongside generator usability.
Main Methods:
- Analysis of 99mTc eluates using gamma spectrometry and thin-layer chromatography.
- Measurement of 99Mo content in the aluminum oxide column.
- Quantification of aluminum in eluates via atomic absorption spectrometry.
- Assessment of elution yield, pH, and practical application of generator accessories.
Main Results:
- All 99mTc eluates exhibited high radionuclidic purity (except one) and excellent radiochemical purity.
- 99Mo breakthrough was not significant, and elution yields ranged from 85% to 122%.
- Eluates consistently met specifications with pH values between 5.0-6.5 and aluminum content below 1 µg/mL.
Conclusions:
- The tested 99Mo/99mTc generators generally performed well, providing a reliable source of 99mTc-pertechnetate.
- The quality control results indicate that these generators are suitable for medical applications, with safe and satisfactory performance.
- Minor exceptions in purity and application did not compromise the overall reliability of the generators.