Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
8.5K
90Sr determination in water samples using Čerenkov radiation
Nataša Todorović1, Ivana Stojković2, Jovana Nikolov1
1University of Novi Sad, Faculty of Sciences, Department of Physics, Novi Sad, Serbia.
Journal of Environmental Radioactivity
|January 31, 2017
Summary
This study optimized a method for detecting strontium-90 (90Sr) in water using Čerenkov radiation. The optimized procedure achieved a low detection limit, crucial for environmental monitoring.
Area of Science:
- Environmental Science
- Radiochemistry
- Analytical Chemistry
Background:
- Accurate determination of radioactive isotopes like strontium-90 (90Sr) in environmental samples is critical for radiological protection.
- Čerenkov radiation coupled with liquid scintillation counting offers a sensitive method for detecting low-activity beta emitters.
- Challenges in Čerenkov counting include color quenching in environmental water samples.
Purpose of the Study:
- To optimize a procedure for determining 90Sr in environmental water samples using Čerenkov radiation and a low-level liquid scintillation counter (Quantulus 1220).
- To evaluate the effect of different vial types on the counting process.
- To apply and validate the sample channel ratio (SCR) method for correcting color quenching.
Main Methods:
- Utilized Čerenkov radiation detection with a Quantulus 1220 liquid scintillation counter.
- Compared the performance of low diffusion polyethylene, high performance glass, and low potassium borosilicate glass vials (20 ml volume).
- Employed the sample channel ratio (SCR) method to correct for color quenching in water samples.
Main Results:
- Achieved counting efficiencies of 45.86(9)%.
- Attained a minimum detectable activity (MDA) of 0.32 Bq L-1 using a 20 ml polyethylene vial over a 300-minute measurement.
- Successfully applied the SCR method to mitigate color quenching effects.
Conclusions:
- The optimized procedure provides a sensitive and reliable method for 90Sr determination in environmental waters.
- Polyethylene vials demonstrated suitability for 90Sr analysis via Čerenkov counting.
- The study's methodology was validated through participation in an international proficiency test (IAEA-TEL-2015-03).