Reliability and validity of non-invasive determined haemoglobin mass and blood volumes
Nazzareno Fagoni1,2, Andreas Breenfeldt Andersen1,3, Laura Oberholzer1
1Zurich Center of Integrative Human Physiology, Institute of Physiology, University of Zürich, Zürich, Switzerland.
Introduction:
The carbon monoxide (CO) rebreathing method used for the determination of haemoglobin mass (Hbmass ) is associated with blood sample analysis (in this study: Radiometer ABL800). As an alternative hereto the aim of the present study was to evaluate the use of a portable and non-invasive CO pulse oximeter (Rad-57).
Method:
With simultaneous determination of CO in the circulation by ABL800 (%HbCO) and Rad-57 (SpCO), Hbmass and blood volume (BV) were determined in duplicates in 24 volunteers. Percentage of typical errors (%TE) within methods and linear correlations between the two procedures were computed.
Results:
Hbmass (Rad-57 = 798 ± 230 g; ABL800 = 781 ± 192 g) and BV (Rad-57 = 5700 ± 1373 ml; ABL800 = 5581 ± 1096 ml) were similar between methods. However, the %TE for Hbmass was higher (P<0·001) for Rad-57 (5·84 ± 5·29%) than for ABL800 (1·35 ± 1·13%). Similarly, the %TE for BV was higher (P<0·001) for Rad-57 (6·06 ± 5·76%) than for ABL800 (1·48 ± 1·25%). Lower (P<0·05) correlation coefficients between the methods were found when Hbmass > 905 g and BV > 6193 ml.
Conclusion:
Assessment of SpCO by Rad-57 resulted in considerably less precise determinations of Hbmass and BV, especially for high values. Thus, non-invasive assessment of Hbmass and BV cannot be recommended for scientific purposes, but may nonetheless be useful in clinical settings.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...