A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Activity Monitors Step Count Accuracy in Community-Dwelling Older Adults
1University of Wisconsin-Eau Claire, USA.
Activity monitors show varied step count accuracy in older adults. Accuracy is generally good for self-selected paces but decreases at slower treadmill speeds, impacting research on walking behaviors.
Area of Science:
- Gerontology
- Biomedical Engineering
- Physical Activity Measurement
Background:
- Activity monitors are widely used to assess physical activity in older adults.
- Accurate step count data is crucial for understanding mobility and health outcomes in this population.
- Previous research has shown mixed results regarding the accuracy of these devices in diverse populations.
Purpose of the Study:
- To evaluate the step count accuracy of various activity monitors in community-dwelling older adults.
- To compare device accuracy across different walking conditions, including self-selected pace and controlled treadmill speeds.
- To inform the selection of appropriate activity monitors for research involving older adult populations.
Main Methods:
- Twenty-nine older adults (mean age ~68 years) participated in the study.
- Step counts from three pedometers and an Actical accelerometer were compared against directly observed steps.
- Data were collected during a 200-m walk and treadmill walking at three distinct speeds (0.67, 0.894, and 1.12 m/s).
Main Results:
- No significant differences were found between activity monitor step counts and actual steps during self-selected pace walking.
- At a slower treadmill speed (0.67 m/s), all devices significantly overestimated or underestimated actual steps.
- Specific devices (Omron HJ-112, Yamax SW-200) showed varying accuracy at different treadmill speeds, with some being accurate at 0.894 m/s but not 1.12 m/s.
Conclusions:
- The choice of activity monitor is critical when studying the walking behaviors of community-dwelling older adults.
- Device accuracy can be compromised at slower walking speeds, potentially affecting research findings.
- Researchers should carefully consider the intended population's typical gait characteristics and walking speeds when selecting activity monitors.