Reliability and Precision of Sonography of the Lumbar Multifidus and Transversus Abdominis During Dynamic Activities
Karl C Gibbon1, Dorothee Debuse1, Angela Hibbs1
1Faculty of Health and Life Sciences, Northumbria University, Newcastle Upon Tyne, England.
Objectives:
To determine the intrarater reliability and precision of lumbar multifidus and transversus abdominis thickness measurements using freehand sonography in a range of static and dynamic conditions.
Methods:
Fifteen asymptomatic participants performed a range of exercises while sonography was used to measure absolute muscle thickness and changes in muscle thickness from rest. Exercise conditions included the abdominal drawing-in maneuver, active straight leg raise, contralateral arm lift, both unloaded and loaded, treadmill walking, and using the Functional Readaptive Exercise Device. Intraday and interday reliability was assessed by intraclass correlation coefficients, and the standard error of measurement was used to assess measurement precision.
Results:
Good to excellent reliability was achieved for absolute transversus abdominis and lumbar multifidus thickness in all conditions. Measurement precision for absolute lumbar multifidus thickness was ≤2.8 mm for the unloaded contralateral arm lift, ≤1.8 mm for the loaded contralateral arm lift, ≤3.1 mm for treadmill walking, and ≤3.8 mm for the Functional Readaptive Exercise Device; for absolute transversus abdominis thickness, precision was ≤0.6 mm for the abdominal drawing-in maneuver, ≤0.5 mm for the active straight leg raise, ≤0.7 mm for treadmill walking, and ≤0.5 mm for the Functional Readaptive Exercise Device. Good to excellent reliability was achieved for relative transversus abdominis and lumbar multifidus thickness in all conditions. Measurement precision for relative lumbar multifidus thickness was ≤3.7% for the unloaded contralateral arm lift, ≤3.8% for the loaded contralateral arm lift, ≤6.3% for treadmill walking, and ≤7.6% for the Functional Readaptive Exercise Device; for relative transversus abdominis thickness, precision was ≤13.6% for the abdominal drawing-in maneuver, ≤6.9% for the active straight leg raise, ≤11.1% for treadmill walking, and ≤7.2% for the Functional Readaptive Exercise Device.
Conclusions:
Acceptable reliability and precision of measurement is achieved for absolute and relative measures of deep spinal muscle thickness using freehand sonography in relatively static and dynamic exercises.
Related Concept Videos
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessment of the Abdomen III: Palpation
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection....