The microINR portable coagulometer: analytical quality and user-friendliness of a PT (INR) point-of-care instrument
Pia Bükmann Larsen1, Elin Storjord2,3, Åsne Bakke4
1a Department of Clinical Biochemistry , Slagelse Hospital , Slagelse, Denmark.
Abstract:
Regular measurement of prothrombin time as an international normalized ratio PT (INR) is mandatory for optimal and safe use of warfarin. Scandinavian evaluation of laboratory equipment for primary health care (SKUP) evaluated the microINR portable coagulometer (microINR®) (iLine Microsystems S.L., Spain) for measurement of PT (INR). Analytical quality and user-friendliness were evaluated under optimal conditions at an accredited hospital laboratory and at two primary health care centres (PHCCs). Patients were recruited at the outpatient clinic of the Laboratory of Medical Biochemistry, St Olav's University Hospital, Trondheim, Norway (n = 98) and from two PHCCs (n = 88). Venous blood samples were analyzed under optimal conditions on the STA-R®Evolution with STA-SPA + reagent (Stago, France) (Owren method), and the results were compared to capillary measurements on the microINR®. The imprecision of the microINR® was 6% (90% CI: 5.3-7.0%) and 6.3% (90% CI: 5.1-8.3) in the outpatient clinic and PHCC2, respectively for INR ≥2.5. The microINR® did not meet the SKUP quality requirement for imprecision ≤5.0%. For INR <2.5 at PHCC2 and at both levels in PHCC1, CV% was ≤5.0. The accuracy fulfilled the SKUP quality goal in both outpatient clinic and PHCCs. User-friendliness of the operation manual was rated as intermediate, defined by SKUP as neutral ratings assessed as neither good nor bad. Operation facilities was rated unsatisfactory, and time factors satisfactory. In conclusion, quality requirements for imprecision were not met. The SKUP criteria for accuracy was fulfilled both at the hospital and at the PHCCs. The user-friendliness was rated intermediate.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Controlled-Current Coulometry: Coulometric Titration
The fundamental requirements for coulometric titrations are (1) 100% efficiency in the reagent-generating electrode reaction and (2) a stoichiometric and preferably rapid reaction between the generated...
Controlled-Current Coulometry: Overview
Automated Microbial Diagnostics