A 2D silicon detector array for quality assurance in small field dosimetry: DUO
Khalsa Al Shukaili1,2, Marco Petasecca1, Matthew Newall1
1Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, 2522, Australia.
The DUO detector, a novel silicon diode array, offers high spatial resolution for precise small field dosimetry in stereotactic radiosurgery/radiotherapy (SRS/SRT). Its real-time capabilities ensure reliable quality assurance for advanced radiation treatments.
Area of Science:
- Medical Physics
- Radiation Oncology
- Detector Technology
Background:
- Small field dosimetry in radiotherapy presents challenges like lateral disequilibrium and source occlusion.
- Accurate quality assurance (QA) is crucial for small field radiotherapy treatments.
- Existing silicon diode arrays often lack the necessary spatial resolution for small field dosimetry.
Purpose of the Study:
- To characterize the DUO monolithic silicon diode array for dosimetry QA in stereotactic radiosurgery/stereotactic radiotherapy (SRS/SRT).
- To evaluate the DUO detector's performance in small radiation fields, focusing on its high spatial resolution (0.2 mm pitch).
Main Methods:
- The DUO detector, comprising two orthogonal 1D silicon arrays with 0.2 mm pitch, was tested.
- Characterization included dose per pulse, percentage depth dose, and spatial resolution measurements.
- Comparisons were made with EBT3 films and ionization chambers for beam profiles and output factors.
Main Results:
- DUO demonstrated dose per pulse stability within 5% across a range of dose rates.
- Percentage depth dose measurements agreed within 1.5% with a Markus ionization chamber.
- Beam profiles showed excellent agreement with EBT3 films (FWHM within 1%, penumbra within 0.5 mm).
- Output factors for small fields agreed within 1.8% with EBT3 films and MOSkin detectors.
Conclusions:
- The DUO monolithic silicon detector array is well-suited for SRS/SRT dosimetry and QA.
- Its high spatial resolution and real-time operation address limitations of previous detectors.
- DUO facilitates accurate measurements in high-gradient radiation fields.