Related Experiment Video
Updated: Mar 8, 2026

Nanoparticle Tracking Analysis of Gold Nanoparticles in Aqueous Media through an Inter-Laboratory Comparison
Published on: October 20, 2020
Interference from Proteins and Surfactants on Particle Size Distributions Measured by Nanoparticle Tracking Analysis
Kelvin Bai1, Gregory V Barnett1, Sambit R Kar1
1Biologics Characterization and Analytical Development, Biologics Development, Bristol-Myers Squibb, 311 Pennington Rocky Hill Road, Pennington, NJ, 08534, USA.
Purpose:
Characterization of submicron protein particles continues to be challenging despite active developments in the field. NTA is a submicron particle enumeration technique, which optically tracks the light scattering signal from suspended particles undergoing Brownian motion. The submicron particle size range NTA can monitor in common protein formulations is not well established. We conducted a comprehensive investigation with several protein formulations along with corresponding placebos using NTA to determine submicron particle size distributions and shed light on potential non-particle origin of size distribution in the range of approximately 50-300 nm.
Methods:
NTA and DLS are performed on polystyrene size standards as well as protein and placebo formulations.
Results:
Protein formulations filtered through a 20 nm filter, with and without polysorbate-80, show NTA particle counts. As such, particle counts above 20 nm are not expected in these solutions. Several other systems including positive and negative controls were studied using NTA and DLS.
Conclusions:
These apparent particles measured by NTA are not observed in DLS measurements and may not correspond to real particles. The intent of this article is to raise awareness about the need to interpret particle counts and size distribution from NTA with caution.
More Related Videos
08:52Improving Reproducibility to Meet Minimal Information for Studies of Extracellular Vesicles 2018 Guidelines in Nanoparticle Tracking Analysis
Published on: November 17, 2021
09:19Nanoparticle Tracking Analysis for the Quantification and Size Determination of Extracellular Vesicles
Published on: March 28, 2021
Related Concept Videos
Factors Affecting Dissolution: Particle Size and Effective Surface Area
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...