Published on: July 7, 2013
Hemodynamics study of a multilayer stent for the treatment of aneurysms
Yan Xiong1, Xuhong Wang2, Wentao Jiang3
1School of Manufacturing Science and Engineering, Sichuan University, 610065, Chengdu, China.
Background:
The changes of hemodynamics caused by the implantation of multilayer stent (MS) have significant effects for aneurysm sac.
Methods:
Comparisons of 3D numerical models with/without a MS in an abdominal aortic aneurysm with a 90° branch vessel were numerically studied from the viewpoint of hemodynamics.
Results:
The results showed that: (1) The flow fields and Wall Shear Stress (WSS) are changed dramatically after MS implantation. The velocity of the blood flow in aneurysm sac decreases significantly and the regions of low-WSS increase. These help thrombus formation; (2) The pressure in aneurysm slightly decreases and keeps close to the normal level of blood pressure, however the risk of aneurysm enlargement or even rupture still exists; (3) The flux and the velocity in branch artery are reduced by about half after MS implantation. Due to the implantation of MS, the changes in the flow field causes the decrease of pressure/WSS in aneurysm sac and the blood flow in branch vessel.
Conclusions:
The implantation of MS into abdominal artery results in more low-WSS regions inside aneurysm which induces thrombus formation. The pressure is reduced slightly means the risk of aneurysm rupture exists.