Perception versus reality-findings from a phosphor plate quality assurance study
Allison Buchanan1, Bryan Benton2, Adam Carraway3
1Associate Professor of Radiology, Department of Oral Health & Diagnostic Sciences, Dental College of Georgia at Augusta University, Augusta, GA, USA.
Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology
|February 5, 2017
Summary
Photostimulable phosphor (PSP) plate image quality remains stable up to 48 clinical uses. Further research is needed to establish optimal replacement guidelines for PSP plates.
Area of Science:
- Radiology
- Medical Imaging Technology
Background:
- Photostimulable phosphor (PSP) plate systems are widely used in digital radiography.
- Assessing the long-term image quality and potential degradation of PSP plates with clinical use is crucial for diagnostic accuracy.
Purpose of the Study:
- To evaluate the image quality of DIGORA Optime PSP plates after repeated clinical use.
- To determine if image degradation occurs and establish potential usage limits.
Main Methods:
- Twenty DIGORA Optime PSP plates were subjected to clinical use and subsequent imaging.
- Image quality was assessed using spatial resolution (line pairs/mm) and contrast resolution via a dental radiographic phantom.
- Contrast-to-noise ratio (CNR) was calculated as a measure of contrast and signal-to-noise.
Main Results:
- Spatial resolution showed a decrease after 48 clinical uses.
- No significant differences were observed in contrast well comparisons.
- Contrast-to-noise ratio (CNR) values showed minimal variation, not exceeding a 0.2 difference.
Conclusions:
- DIGORA Optime PSP plates maintain unaffected image quality for up to 48 clinical uses.
- Additional studies are recommended to validate these findings and develop guidelines for PSP plate replacement intervals.
Related Concept Videos
Photoluminescence: Applications
1.2K
Photoluminescence offers a wide range of applications due to its inherent sensitivity and selectivity. This technique allows for both direct and indirect analyses of the analyte. Direct quantitative analysis is possible when the analyte exhibits a favorable quantum yield for fluorescence or phosphorescence. However, an indirect analysis may be feasible if the analyte is not fluorescent or phosphorescent, or if the quantum yield is unfavorable. Indirect methods include reacting the analyte with...
1.2K
Fluorescence and Phosphorescence: Instrumentation
1.7K
Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions.