Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
Survival of the pegged glenoid component in shoulder arthroplasty: part II
Paul B McLendon1, Bradley S Schoch1, John W Sperling1
1Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN, USA.
Background:
Loosening of the glenoid component is a primary reason for failure of an anatomic shoulder arthroplasty. Pegged glenoids were designed in an effort to outperform keeled components. This study evaluated the midterm clinical and radiographic survival of a single implant design with implantation of an in-line pegged glenoid component and identified risk factors for radiographic loosening and clinical failure.
Materials And Methods:
There were 330 total shoulder arthroplasties that had been implanted with a cemented, all-polyethylene, in-line pegged glenoid component evaluated with an average clinical follow-up of 7.2 years. Of these shoulders, 287 had presurgical, initial postsurgical, and late postsurgical radiographs (mean radiographic follow-up, 7.0 years).
Results:
At most recent follow-up, 30 glenoid components had been revised for aseptic loosening. This translated to a rate of glenoid component survival free from revision for all 330 shoulders of 99% at 5 years and 83% at 10 years. Of 287 glenoid components, 120 were considered loose on the basis of radiographic evaluation. Four humeral components were considered loose. Component survival (Kaplan-Meier) free from radiographic failure at 5 and 10 years was 92% and 43%. Severe presurgical glenoid erosion (Walch A2, B2, C) and patient age <65 years were risk factors for radiographic failure. Late humeral head subluxation was associated with radiographic failure.
Conclusion:
Despite the predominant thinking that pegged glenoid components may be superior to keeled designs, midterm radiographic and clinical failure rates were high with this pegged component design, particularly after 5 years. Advanced presurgical glenoid erosion and younger patient age are risk factors for radiographic loosening. Revision rates underestimate radiographic glenoid loosening.