A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Carrying loads: Validating a portable tri-axial accelerometer during frequent and brief physical activity
G Zorrilla-Revilla1, A Mateos2, O Prado-Nóvoa2
1Escuela Interuniversitaria de Posgrado en Evolución Humana, Universidad de Burgos, Spain.
Objectives:
The aim of this study is to evaluate the validity of the SenseWear Armband™ (SWA, Model MF-SW, BodyMedia Inc.) as a tool to assess daily brief, light- to moderate-intensity activities.
Design:
A total of 41 volunteers were recruited (27 males, 14 females) to perform several trials, including a resting metabolic rate test and a number of walking trials while carrying different loads in a backpack.
Methods:
Energy expenditure during trials was measured using both the SWA™ mini device, with the indirect calorimetry method (MasterScreen CPX and Oxycon Mobile JAEGER™ devices) used for validated comparative measurement.
Results:
The SWA™ mini shows agreement with indirect calorimetry in all trials. However, the SWA™ mini over-estimated expenditure in all participants. Individual assessment estimates with the SWA™ mini also exhibited random errors. The variations in energy expenditure (EE) resulting from increased carried loads during the trials were not statistically significant when EE was measured with the SWA™ mini. Furthermore, Metabolic Equivalent of Task (MET) calculation, highly dependent on estimated energy expenditure per unit time, also was likely overestimated. In contrast, the SWA™ mini provided estimates of the resting metabolic rate with a small error.
Conclusions:
The SWA™ mini is not a valid device for estimating energy expenditure in brief light- or moderate activities.
Related Concept Videos
Measuring Acceleration Due to Gravity
A simple pendulum can be described as a point mass and a string. Meanwhile, a physical pendulum is any object whose oscillations are similar to a simple pendulum, but cannot be modeled as a point mass on a string because its mass is distributed over a larger area. The behavior of a physical pendulum can be modeled using the principles of...