Related Experiment Video
Updated: Mar 8, 2026

Monitoring the Wall Mechanics During Stent Deployment in a Vessel
Published on: May 8, 2012
Longitudinal deformation of a third generation zotarolimus eluting stent: "The concertina returns!"
Vasileios F Panoulas1, Ozan M Demir1, Neil Ruparelia1
1Vasileios F Panoulas, Ozan M Demir, Neil Ruparelia, Iqbal Malik, Cardiology Department, Imperial College Healthcare NHS Trust, Hammersmith Hospital, London W120HS, United Kingdom.
Abstract:
In the current case series we describe two cases of longitudinal stent deformation in ostial lesions treated with a new generation zotarolimus eluting stent and review current literature on longitudinal stent deformation. Historically not a common occurrence, longitudinal deformation occurred mainly in Promus Element everolimus eluting stents, which had only two rather than the commonly used 3 links between stent rings. Longitudinal deformation commonly occurs secondary to compression of the proximal edge of the stent by either the guide catheters, or intravascular balloons and imaging catheters. The degree of deformation however, depends on the longitudinal strength and design of the stent.